Pour PlateRepresenting 150 bacteria
SimPlate® for HPCRepresenting 150 bacteria (under UV light)
SimPlate® for HPC is US EPA-approved.
Internal comparisons of SimPlate® for HPC and Standard Methods (9215B) pour plate method show a correlation coefficient (r) of 0.97.
Step 1. Add 100 mL sterile diluent to media vessel.
Step 2.Add 1 mL sample to SimPlate.
Step 3.Add 9 mL media to SimPlate.
Step 4.Swirl SimPlate and pour off excess.
Step 5.Invert and incubate for 48 hours.
Step 6. Read results under UV light.
Back to top
SIMPLATE® is the registered trademark of BioControl Systems, Inc. and is used by IDEXX under license from BioControl. Covered by U.S. Patent No. 5,700,655. Other patents pending.