IDEXX > Companion Animal > In-House Tests > SNAP 4Dx Test > How to Use
 
 
Canine SNAP 4Dx Test

It's very easy to screen for infection.

 

Blood sample test procedure:

  1. Mix the sample and conjugate in a sample tube.

  2. Add the contents of the sample tube to the sample well of the SNAP device.

  3. When color first appears in the activation circle, press the activator to activate the device.

  4. Read the test result.


test procedure
test procedure


results


Important notes on storage:

  • Room-temperature storage (optional): SNAP devices and reagents can be stored at room temperature (15°–27°C/59°–80°F) for 90 days or until the printed expiration date, whichever occurs first.

  • After SNAP devices and reagents are removed from 2°–7°C (35°–45°F) storage for more than 24 hours, the expiration date is 90 days or the printed expiration date, whichever occurs first.

  • If the 90-day expiration date occurs prior to the printed expiration date, record the new date in the space provided on the kit.

  • All components must be at room temperature before running the test. This may take 30 minutes, depending on the temperature of your laboratory—do not heat.

For more information, contact your authorized IDEXX distributor or call 1-800-248-2483.

View the complete SNAP 4Dx Test package insert. (602 KB)