When testing with the NSAID Monitoring Panel, we recommend using a lithium-heparin tube with a gel barrier.

Plasma
Separate plasma from red cells within five minutes when using lithium-heparin.

 
Download Our Poster
Ensure maximum accuracy—always prepare the sample properly when analyzing blood chemistry parameters. (1.1 KB)
 
  1. preparation graphicsUse the appropriate tube. DO NOT USE EDTA.
  2. Use the appropriate sample collection device.
  3. centrifugeDraw the sample gently. Transfer if necessary. Use the correct blood-to-lithium-heparin ratio.
  4. Gently invert the sample for 30 seconds to mix.
  5. Centrifuge the samples.
    Standard Centrifuge
    (8,000 rpm): 10 minutes

    sample cup transferHigh-Speed Centrifuge
    (StatSpin® ≥ 12,000 rpm): 120 seconds

    StatSpin Settings:
    StatSpin MP—Hematocrit Setting
    StatSpin VT—Hard Spin
  6. Transfer the plasma sample to a sample cup.
 