| Parameters |
|
Hct (%), Hgb (g/dL), MCHC (g/dL), total WBC (#), GRANS (% and absolute), LYMPHS/MONOS (% and absolute), PLT (#) and fibrinogen* (mg/dL). In canine and bovine samples only, the analyzer will also delimit neutrophils (absolute) and eosinophils (absolute). In canine and feline samples only, the analyzer will also delimit reticulocytes (%). |
|
| Units |
|
Conventional or SI units |
|
| Sample Size |
|
111 µL whole blood with EDTA anticoagulant is used with each IDEXX VetTube™ per evaluation. |
|
| Result Timing |
|
Prepared IDEXX VetTubes™ are centrifuged for five minutes and then placed in the IDEXX VetAutoread analyzer for approximately one minute to completely process a patient specimen, report complete results and create a buffy coat profile. To precipitate plasma fibrinogen, a previously centrifuged IDEXX VetTube™ is placed in the IDEXX Fibrinogen Precipitator for five minutes, centrifuged for five additional minutes, and placed in the IDEXX VetAutoread analyzer for approximately one minute for results. |
|
| Calibration |
|
Available with the IDEXX VetAutoread calibration (cal) rod |
|
| Power Supply |
|
Universal power supply |
|
| Operating Temperature |
|
IDEXX VetAutoread: 20°32°C (68°90°F)
IDEXX VetCentrifuge™: 20°32°C (68°90°F)
IDEXX Fibrinogen Precipitator: 20°32°C (68°90°F) |
|
| Dimensions |
|
IDEXX VetAutoread: 34.3W x 24.13D x 10.2H cm
IDEXX VetCentrifuge™ (lid closed): 23.5 x 29.8 x 12.7 cm
IDEXX Fibrinogen Precipitator: 11.43 x 13.97 x 4.40 cm |
|
| Net Weight |
|
IDEXX VetAutoread: 3.6 kg (8 lb)
IDEXX VetCentrifuge™: 2.8 kg (6.2 lb) |
|
| Accessories |
|
Power pack, pipettor, workstation, accessory pack, calibration rod, tweezers, screwdriver
| IDEXX VetCentrifuge™ |
Starter Kit |
Power pack
Power cords
Accessories pack
Rotor wrench
Rotor removal wire |
Software cartridge
Operator's guide
IDEXX VetTubes™ (one tray)
VetTest® connection cable |
View the IDEXX VetLab® system accessories. |