| Dimensions and Weight |
|
Depth: 9.4 cm (3.7 in)
Width: 19 cm (7.5 in)
Height: 5 cm (2.0 in)
Weight: 0.53 kg (1.2 lbs) |
|
| Performance |
|
Test Precision: ≤10% C.V. for whole blood samples |
|
| Operation |
|
Test Chamber: 1
Timing Range: 0 seconds to 500 seconds
Incubation Temperature: 37°C ± 1.0°C (98.6°F ± 2.0°F)
Full-Charge Operating Time: >2 hours
Battery Life: 500 recharges
Batterty Type: Lithium Ion
Throughput (Full Charge): 49 test cycles (at 150 sec per test), 17 test cycles (>500 sec per test)
Operating Environment: 15°C to 30°C (59°F to 86°F) |
|
| AC/DC Power Module |
|
Input Power: 100 to 240 VAC, 50 to 60 Hz
Output Power: 12.0 VDC, 3.5 Amps (42 Watts, 144 BTU/hr) minimum |
|
| Calibration |
|
The IDEXX Coag Dx Analyzer is calibrated at the manufacturing
facility to test and verify all functions. The analyzer is also self-calibrating,
as all analyzer functions are continuously monitored
and verified by the analyzer software when a test is performed.
The analyzer does not require additional calibration by the user. |